How do you install the Fiberglass Lab Blowers?

Fiberglass Lab Blower
Installation and Operating Instructions

Part Number: 37200 and 37201

 

Your new fiberglass blower comes fully assembled and ready for installation. Securely mount the blower to s flat surface with bolts or leg screws. The blower is mounted in the “up-blast” position. This position can be changed for your application by loosening the two bolts and nuts which hold the blower housing clamp to the blower housing base. Rotate the housing to the desired position and e-tighten the bolts and nuts.

The blower with the standard 1/3 Hp motor comes with a cord attached ready to be plugged into a standard 110 volt wall outlet.

The blower with the explosion proof motor must be wired according to local electrical codes for explosion proof Class D motors. (See your local electrician)

The inlet and the outlet of the blower are 8” In diameter so that 8” flexible tubing can be clamped to either or both of the inlet and outlet. If you are installing rigid PVC duct to the blower, use the supplied pieces of pressure sensitive closed cell foam to insure a tight fit between the PVC duct and the bower inlet or outlet.

Periodically check the blower wheel to insure that it is tight on the shaft and that there are no cracks in the polypropylene blower wheel. Check the motor to insure that it is fastened tightly to the motor on the base.

Please contact us for replacement parts.

No manual or diagrams are available for this blower.

Pre-wire plug in on the Open Drip Proof Motor runs counter clockwise (sleeve bearing) 180 degrees.

Blowers have stainless steel shaft, fiberglass housing, and polypropylene wheel.

Stainless Steel Extender shaft is 5/8” female X 1/2” male

 

Motor Info

Explosion-proof Motor: Hazardous location, Capacitor Start Totally Enclosed Fan Cooled Motor, HP 1/3, RPM 1725, Voltage 115/230 V, NEMA Frame 56, Service Factor1.00, Frequency 60 HZ, Mounting Rigid, Bearings Double Shielded Ball, Thermal Protection Automatic, Full Load Amps 7.6/3.8, Meets NEC Hazardous Atmosphere Class 1 Group C and D And Class 2 Groups E, F, and G, Rotation CW/CCW, Shaft Diameter 5/8”,  Shaft Length: 1-7/8”

 

Open Drip-proof Motor: Belt Drive Motor, Split Phase, Open Drip-proof, 1/3 HP, 1725 RPM, Number of speeds 1, 115 Volt, 6.6 Full load Amps, 56 NEMA Frame, Service Factor 1.35, 60 Hz, Cradle Mount. Sleeve Bearings, Auto Thermal Protection. Rotation CW/CCW, Ambient 40° C, Insulation Class B, Shaft Diameter 5/8”, Shaft Length 1-7/8”, Body Diameter 5-5/8”

Article Title:
How do you install the Fiberglass Lab Blowers?
Article Summary:
Installation and Operating Instructions for the Fiberglass Lab Blowers.
Article Date:
5/29/2008
Article ID:
429
0 user(s) found this article useful. Was this article useful to you?
Share

Information disclaimer:

The information contained in this article is for general information purposes only. The information is provided by United States Plastic Corp. and while we endeavor to keep the information up to date and correct, we make no representations or warranties of any kind, express or implied, about the completeness, accuracy, reliability, suitability or availability with respect to the website or the information, products, services, or related graphics contained on the website for any purpose. Any reliance you place on such information is therefore strictly at your own risk.

In no event will we be liable for any loss or damage including without limitation, indirect or consequential loss or damage, or any loss or damage whatsoever arising from loss of materials, time or profits arising out of, or in connection with, the use of this information.